CORONARY PHYSIOLOGY: SmartMap® Pressure Instrument

Simple Is Better
Easy to use with plug-and-go set up. Automatic calibration.

Pressure Without Pressure!

So Portable It Can Go Anywhere!
  • Small size means convenient set up in any cath lab.
  • System so versatile it can be mounted on I.V. pole or bedside, or bagged in sterile field.

TECHNICAL SPECIFICATIONS
System:
  • Pressure Measurement Range: -30 to +330 mmHg
  • Pressure Band Width (@-3 db): DC to 25 Hz
  • Pressure Measurement Delay Time: <40 ms
  • Measurement Accuracy: ±3 mmHg (-30 to + 100 mmHg) ±3% (>100 mmHg)
  • Analog Output: 5µV/V/mmHg (adjustable between 4-6µV/V/mmHg)
Power Requirements:
  • Vin DC Min 2.4 V
  • Max 11 V
  • Vin AC, 1 KHz to 5 KHz sine Min 2.5 Vrms
  • Max 8 Vrms
  • Vin DC, 500Hz to 1 KHz sine Min 3 Vrms
  • Max 8 Vrms
AC Frequency Excitation:
  • 500 Hz to 5 KHz sine wave
  • 0 to 5 KHz;
  • rise/fall time < 10 µs
Vex Current:
  • <10 mA@4 to 11 Vrms
  • <19 mA below 4Vrms
Vex Current Limit:
  • 25mA max ±3 mA
Dimensions:
  • 15.75 cm W x 8.853 cm H x 4.48 cm / 6.3" W x 3.54" H x 1.79" D
Weight:
  • 0.22 Kg or 220 grams / 0.45 pound

ORDER INFORMATION

Model 6500 SmartMap® Pressure Instrument
For more information on Volcano Products, please click here.

For product IFU files, please click here.